Please use this identifier to cite or link to this item:
https://hdl.handle.net/2440/60686
Citations | ||
Scopus | Web of Science® | Altmetric |
---|---|---|
?
|
?
|
Type: | Journal article |
Title: | EZYprep LC-coupled MALDI-TOF/TOF MS: an improved matrix spray application for phosphopeptide characterisation |
Author: | Condina, M. Gustafsson, O. Klingler-Hoffmann, M. Bagley, C. McColl, S. Hoffmann, P. |
Citation: | Proteomics, 2010; 10(13):2516-2530 |
Publisher: | Wiley - V C H Verlag GmbH & Co. KGaA |
Issue Date: | 2010 |
ISSN: | 1615-9853 1615-9861 |
Statement of Responsibility: | Mark R. Condina, Johan O. R. Gustafsson, Manuela Klingler-Hoffmann, Christopher J. Bagley, Shaun R. McColl and Peter Hoffmann |
Abstract: | The quality of MALDI-TOF mass spectrometric analysis is highly dependent on the matrix and its deposition strategy. Although different matrix-deposition methods have specific advantages, one major problem in the field of proteomics, particularly with respect to quantitation, is reproducibility between users or laboratories. Compounding this is the varying crystal homogeneity of matrices depending on the deposition strategy used. Here, we describe a novel optimised matrix-deposition strategy for LC-MALDI-TOF/TOF MS using an automated instrument that produces a nebulised matrix “mist” under controlled atmospheric conditions. Comparisons of this with previously reported strategies showed the method to be advantageous for the atypical matrix, 2,5-DHB, and improved phosphopeptide ionisation when compared with deposition strategies for CHCA. This optimised DHB matrix-deposition strategy with LC-MALDI-TOF/TOF MS, termed EZYprep LC, was subsequently optimised for phosphoproteome analysis and compared to LC-ESI-IT-MS and a previously reported approach for phosphotyrosine identification and characterisation. These methods were used to map phosphorylation on epidermal growth factor-stimulated epidermal growth factor receptor to gauge the sensitivity of the proposed method. EZYprep DHB LC-MALDI-TOF/TOF MS was able to identify more phosphopeptides and characterise more phosphorylation sites than the other two proteomic strategies, thus proving to be a sensitive approach for phosphoproteome analysis. |
Keywords: | ESI-IT-MS/MS LC MALDI matrix preparation phosphorylation technology |
Rights: | © 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim |
DOI: | 10.1002/pmic.200900800 |
Published version: | http://dx.doi.org/10.1002/pmic.200900800 |
Appears in Collections: | Aurora harvest IPAS publications Molecular and Biomedical Science publications |
Files in This Item: